The sensor consists of transmitter part and display output part, the transmitter part is mainly composed of detecting element, signal amplifier circuit and microcontroller, which completes the detection of ambient differential pressure, adiabatic pressure, temperature and humidity, and transmits the measured value to the display output part, which carries out the digital display and external signal output and realizes the acousto-optic alarm output in comparison with the alarm point; the detecting element adopts the high-precision sensor element to realize the detection of differential pressure, adiabatic pressure, temperature and humidity. The detecting element adopts high-precision sensor elements to realize the measurement of differential pressure, absolute pressure, temperature and humidity.
High Integration: The integration of four measurement functions into a single unit significantly reduces the number of underground monitoring devices required. This not only conserves valuable installation space but also substantially lowers deployment and maintenance costs.
① Remote Adjustment Function
Supports remote operation within a 5-meter range, with an effective control angle of 0° to 30°, enabling flexible and convenient on-site debugging.
② Multi-Parameter Independent Alarm Setting
Allows individual alarm point configuration for each of the four parameters. Users can freely set alarm thresholds within the measurable range of each parameter.
③ RS485 Communication Function
Equipped with an RS485 interface with configurable baud rate. The default factory setting is 9600 bps, and the communication mode is fixed at even parity.
④ Multi-Alarm Output Mechanism
When any parameter exceeds its limit, the sensor immediately triggers both audible and visual alarms. Simultaneously, the switch output changes from low to high level.
l Audible alarm intensity: ≥80 dB(A) at 1 meter
l Visual alarm visibility: Clearly visible from 20 meters in dark conditions
Operating voltage range: DC 9 ~ 24.5V
Rated operating voltage: DC 18V
Maximum working current: ≤100mA
② Measuring Range and Error
Differential pressure:
‒500~500Pa
Error
≤±2Pa
Resolution
1Pa
Absolute pressure:
60~110kPa
Error
≤±30Pa
Resolution
1Pa
Temperature:
-10℃~80℃
Error
≤±1℃
Resolution
0.01℃
Humidity:
0~100%RH
Error
≤±3%RH
Resolution
0.024%RH
Differential pressure: -500~500Pa, Error ≤±2Pa, Resolution 1Pa.
Absolute pressure: 60 ~ 110kPa, Error ≤ ± 30Pa, Resolution 1Pa
Temperature: -10℃~80℃, Error≤±1℃, Resolution 0.01℃.
Humidity: 0~100%RH, error≤±3%RH, resolution 0.024%RH
③ Response characteristics
Temperature response time: minimum 5s
④ Output signal
RS485 output
Interface number: 1 channel
Transmission mode: RS485, master-slave, half-duplex
Transmission rate: 1200~38400bps
Transmission distance: ≤2000m (when using 1.5mm² wire cross-section)
Signal voltage peak: 2~12V
⑤ Switching output
Interface number: 1 channel
Output voltage: DC 24V
Maximum current: 100mA
Functional features
Technical parameters
① Power Supply Performance